Feature #1285
Updated by tin over 6 years ago
**Computation of the drift rate of calibration constant 72 (will take 1 week to complete):
C = [(A – B) * 1000000] / [A * D] ppm per day
A = the value of the calibration constant 72 (following an ACAL on day 1 of the test)
B = the value of calibration constant 72 (following an ACAL on the last day of the test)
C = the drift rate of calibration constant 72 in ppm per day
D = the number of test days where the instrument was powered on during the test interval
(See Note # 1)
If the absolute value of C is > 0.43 ppm per day, then the A3 assembly in your instrument needs to be
replaced.
*Complete an ACAL just prior to obtaining the present value of Calibration Constant 72*
|Date |CAL 72 after ACAL|ppm Deviation|TEMP?|
|11/16| 981.648769E-3 |Initial| +41.3°C |
|11/16 +3h| 981.652110E-3 |+0.334| +35.5°C |
|11/17| | |
|11/18| | |
|11/19| | |
|11/20| | |
|11/21| | |
|11/22| | |
|11/23| | |
|11/24| | |
|11/25| | |
|11/26| | |