Description

Using Keithley 2001 K7 unit
Room temp: 24.3°C (HEL-700 RTD set to 999.999 ohm on same Keithley unit)
20V Range : ±18ppm + 8 counts (Keithley 2001) + absolute ± 100ppm from Tektronix Cal lab.
10V : ±18ppm + 8 counts (Keithley 2001) + absolute ± 54ppm from Tektronix Cal lab.
2V Range : ±18ppm + 4 counts (Keithley 2001) + absolute ± 13ppm from Tektronix Cal lab.
200mV Range : ±25ppm + 12 counts (Keithley 2001) + absolute ± 2.1ppm from Tektronix Cal lab.

8/FEB/2014

Initial cal

1. Zener voltage 1557 - adjustment R1 to set zener voltage. Must be within 100uV

Temperature Zener voltage Vz Zener current Iz D ppm Meas uncertainty
Label 23 to 25C 6.1698 V 7.5 mA ?
Measured 24.3 C 6.169113 V Not measured -111.3 ±18ppm + 8 counts & absolute ± 100ppm
Adjusted 24.3 C 6.169801 V Not measured 0.16 ±18ppm + 8 counts & absolute ± 100ppm

2. Adjust R2 for zero offset, all decades set to zero, DVM on output terminals (local sense)

Measured Vos 24.3 C -0.01629 mV
Adjusted 24.3 C -0.00203 mV

2A. Adjust R2a for sense "0" offset. Set all decades to zero, polarity to plus, range 10V.
Remove sense link from Black (LO) terminals and measure black terminals. Adjust to be less than 1 mV

Measured Vszo 24.3 C 0.81512 mV
Adjusted 24.3 C -0.00101 mV

3. Adjust R3 to set feedback current accurately as required for accurate Vout

Dial 1V on second decade, using 10 position. DVM on output terminals, output read as 1.0000 +- 50uV

Measured Voi 24.3 C 0.9998574 -142.6 ppm
Adjusted 24.3 C 1.0000004 V 0.4 ppm

4. Linearity adjustment of 1st decade

Set setting same as step 3. Set 2nd decade to zero.

5. Set 2V on 1st decade. Adjust P2 pot to output. Set 1V on 1st decade, adjust P1 to 1V.
Vc - after initial cal settings done.

Measured 2V 24.3 C 1.9997686 -115,7 ppm
Measured 2Vc 24.5 C 2.0000546 27,3 ppm
Adjusted 24.5 C 1.9999995 0,3 ppm
Measured 1V 24.5 C 0.9998835 -116,5 ppm
Measured 1Vc 24.5 C 1.0000144 14,4 ppm
Adjusted 24.5 C 1.0000001 0,1 ppm

6. Set 4V on 1st decade to adjust P4. Set 3V to tweak P4.

Measured 4V 24.5 C 3.999532 -117 ppm
Measured 4Vc 24.5 C 4.000046 11,5 ppm
Adjusted 24.5 C 4.000003 0,75 ppm
Measured 3V 24.5 C 2.999645 -118.3 ppm
Measured 3Vc 24.5 C 3.000038 12.7 ppm
Adjusted 24.5 C 2.999994 -2 ppm

7. Set 6V on 1st decade to adjust P6. Set 5V to tweak P6.

Measured 6V 24.4 C 5.999280 -120 ppm
Measured 6Vc 24.5 C 6.000030 5 ppm
Adjusted 24.4 C 6.000007 1,2 ppm
Measured 5V 24.4 C 4.999398 -120 ppm
Measured 5Vc 24.5 C 5.000027 5,4 ppm
Adjusted 24.4 C 5.000003 0,6 ppm

8. Set 8V on 1st decade to adjust P8. Set 7V to tweak P8.

Measured 8V 24.4 C 7.999022 -122,3 ppm
Measured 8Vc 24.5 C 8.000021 2,6 ppm
Adjusted 24.4 C 8.000002 0,3 ppm
Measured 7V 24.4 C 6.999138 -123,1 ppm
Measured 7Vc 24.5 C 7.000018 2,5 ppm
Adjusted 24.4 C 7.000001 0,2 ppm

9. Set 10V on 1st decade to adjust P10. Set 9V to tweak P10.

Measured 10V 24.4 C 9.998775 -122,5 ppm
Measured 10Vc 24.6 C 10.000050 5 ppm
Adjusted 24.6 C 10.000002 0,2 ppm
Measured 9V 24.4 C 8.998888 -123,5 ppm
Measured 9Vc 24.6 C 9.000053 5,9 ppm
Adjusted 24.6 C 9.000001 0,1 ppm

Range set to 100mV

Set setting same as step 3. Set decades to zero.

5. Set 20mV on 1st decade. Adjust P4 pot to output.

Measured 20mV 24.4 C 19.99800 -100 ppm
Measured 20mVc 24.7 C 19.99994 -3 ppm
Adjusted 24.7 C 19.99998 -1 ppm

6. Set 40mV on 1st decade to check

Measured 40mV 24.4 C 39.99562
Adjusted 24.7 C 39.99952

7. Set 60mV on 1st decade to check

Measured 60mV 24.4 C 59.99295
Adjusted 24.7 C 59.99938

8. Set 80mV on 1st decade to check

Measured 80mV 24.4 C 79.99011
Adjusted 24.7 C 79.99916

9. Set 100mV on 1st decade to check

Measured 100mV 24.4 C 99.98775 -122,5 ppm
Adjusted 24.7 C 99.99878 -12,2 ppm

Range set to 10mV

5. Set 2mV on 1st decade.

Measured 2mV 24.4 C 1.99891 -545 ppm
Adjusted 24.7 C 1.99926 -370 ppm

6. Set 4mV on 1st decade to check

Measured 4mV 24.4 C 3.99777
Adjusted 24.7 C 3.99837

7. Set 6mV on 1st decade to check

Measured 6mV 24.4 C 5.99670
Adjusted 24.7 C 5.99758

8. Set 8mV on 1st decade to check

Measured 8mV 24.4 C 7.99581
Adjusted 24.7 C 7.99631

9. Set 10mV on 1st decade to check

Measured 10mV 24.4 C 9.99467 -533 ppm
Adjusted 24.7 C 9.99544 -456 ppm

Measured means before calibration
Adjusted - after calibration

#1

Updated by tin about 9 years ago

  • Description updated (diff)
#2

Updated by tin about 9 years ago

  • Description updated (diff)
#3

Updated by tin about 9 years ago

  • Description updated (diff)
#4

Updated by tin about 9 years ago

  • Description updated (diff)
#5

Updated by tin about 9 years ago

  • Description updated (diff)
#6

Updated by tin about 9 years ago

Perf log, 4-wire connect

KI2001 Cal, 8.FEB.14 KI2002 2TC, 21.2, 16.MAR14 KI2002,2TC,24.8,18.MAR14
10Vrng, 0V -0.00203 mV 2030 0.000477 mV 477 0.000225 mV 225
10Vrng, 1V 1.0000001 V 0.1 1.00000936 V 9.36 1.00000333 V 3.33
10Vrng, 2V 1.9999995 V 0.3 2.00001817 V 9.1 2.00000408 V 2.04
10Vrng, 3V 2.999994 V -2 3.0000062 V 2 2.9999936 V -2.1
10Vrng, 4V 4.000003 V 0.75 4.0000161 V 4 3.9999956 V -1.1
10Vrng, 5V 5.000003 V 0.6 5.0000156 V 3.12 5.0000017 V 0.34
10Vrng, 6V 6.000007 V 1.2 6.0000224 V 3.73 6.0000054 V 0.9
10Vrng, 7V 7.000001 V 0.2 7.0000183 V 2.61 7.0000026 V 0.37
10Vrng, 8V 8.000002 V 0.3 8.0000277 V 3.46 8.0000032 V 0.4
10Vrng, 9V 9.000001 V 0.1 9.0000222 V 2.47 8.9999964 V -0.4
10Vrng,10V 10.000002 V 0.2 10.0000374 V 3.74 9.9999994 V 0.06