Project

General

Profile

Feature #895

Feature #891: Self-test errors diagnosis

304.x series self-test errors (Ohm)

Added by tin about 9 years ago. Updated over 7 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Target version:
Start date:
01/03/2014
Due date:
01/05/2014
% Done:

0%

Estimated time:

Description

+8VF and -8VF have incorrect signal


Files

compar-refok001.png (6.88 KB) compar-refok001.png PASS WFM -8VF tin, 01/03/2014 03:11 AM
compar-refok002.png (6.7 KB) compar-refok002.png PASS WFM +8VF tin, 01/03/2014 03:11 AM
compar-refok003.png (6.75 KB) compar-refok003.png FAIL WFM -8VF tin, 01/03/2014 03:11 AM
compar-refok004.png (6.74 KB) compar-refok004.png FAIL WFM +8VF tin, 01/03/2014 03:11 AM
Keithley_2001J _ 03 _ FLOAT PSU.pdf (123 KB) Keithley_2001J _ 03 _ FLOAT PSU.pdf tin, 01/03/2014 03:32 AM
top_rev_04_JAN_14.jpg (6.36 MB) top_rev_04_JAN_14.jpg Board coverage. tin, 01/04/2014 03:09 AM
#2

Updated by tin about 9 years ago

По схеме на данный момент

Проверить участок

#3

Updated by tin over 8 years ago

  • Target version changed from Keithley 2001 #4 to Keithley 2001 #5
#4

Updated by Izerg about 8 years ago

  • Status changed from In Progress to Suspended
#5

Updated by tin over 7 years ago

  • Status changed from Suspended to Closed

Also available in: Atom PDF