Feature #885
Keithley 2001 thermal evaluation
Start date:
01/01/2014
Due date:
01/01/2014
% Done:
40%
Estimated time:
4.00 h
Description
Thermal camera : Fluke Ti32
Files
Updated by tin about 9 years ago
- Due date changed from 12/29/2013 to 01/01/2014
- Start date changed from 12/29/2013 to 01/01/2014
Postponed to 1/1/14
Updated by tin about 9 years ago
- Category set to 60
- Status changed from New to In Progress
- Target version set to Keithley 2001 #4
- % Done changed from 0 to 40
Processing data.
Captured thermal images for
Keithley unit #4
Analog board: A09891
Second analog board from Keithley #1 (passes selftest on 110VAC after warmup)
ADC board: A39577
Updated by tin about 9 years ago
- File k1_top1.jpg k1_top1.jpg added
- File k1_top2.jpg k1_top2.jpg added
- File a09891_top1.jpg a09891_top1.jpg added
- File a09891_top2.jpg a09891_top2.jpg added
- File a09891_R286.jpg a09891_R286.jpg added
- File k1_R286.jpg k1_R286.jpg added
- File k1_right_a98_left_U324.jpg k1_right_a98_left_U324.jpg added
Added exported images
Keithley #1 running self-test continuously , PASS
Keithley #4 running self-test continuously , FAIL
Suspected failures: R286 related components, CR335, U324 related components, Q540
Updated by tin about 9 years ago
- File a09891_vref3.jpg a09891_vref3.jpg added
- File a09891_vref2.jpg a09891_vref2.jpg added
- File a09891_vref.jpg a09891_vref.jpg added
- File IR000091.png IR000091.png added
- File IR000089.png IR000089.png added
- File IR000088.png IR000088.png added
- File IR000093.png IR000093.png added
- File IR000092.png IR000092.png added
- File IR000075.jpg IR000075.jpg added
- File a09891_back.jpg a09891_back.jpg added
Updated by tin about 9 years ago
- File a09891_adc_off.jpg a09891_adc_off.jpg added
- File adc_top_off.jpg adc_top_off.jpg added
- File OR000167.jpg OR000167.jpg added