- Status changed from New to In Progress
AC input test, self-test result
AC source: Chroma 61504
110VAC, 60Hz - PASS
230VAC, 50Hz - FAIL2
90VAC | 132VAC | 180VAC | 250VAC | |
50Hz | FAIL2 | FAIL2 | FAIL2 | FAIL2 |
60Hz | PASS | PASS | PASS | PASS |
400Hz | FAIL1 | FAIL2 | FAIL1 | FAIL2 |
FAIL1 - Errors 200.1,200.2,200.6,200.7,201.2,401.1
FAIL2 - Errors 200.1,200.6,200.7,201.2
EDC MV106 1.00000V with REL set measurement overnight
Vmin = 1.0000001VDC, Vmax 1.0000042VDC
Updated by tin almost 9 years ago
- Status changed from In Progress to Closed
- % Done changed from 0 to 100
Calibrated by Tek/Kei
Standards used:
F5700A + 5725A, Kei 2001-758 1Gohm standard, Agilent 33220A
Updated by tin almost 9 years ago
AC input test, self-test result after cal
AC source: Chroma 61504
110VAC, 60Hz - PASS
230VAC, 50Hz - FAIL2
90VAC | 132VAC | 180VAC | 250VAC | |
50Hz | PASS | PASS | PASS | PASS |
60Hz | PASS | PASS | PASS | PASS |
400Hz | FAIL3 | PASS | FAIL3 | PASS |
FAIL3 - Errors 200.2,200.3,401.1
EDC MV106 1.00000V with REL set measurement overnight
Updated by tin almost 8 years ago
- File ACK7-2.jpg ACK7-2.jpg added
- File ACK7-3.jpg ACK7-3.jpg added
- File ACK7-5.jpg ACK7-5.jpg added