Project

General

Profile

Feature #869

Spec test

Added by tin almost 6 years ago. Updated over 4 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Target version:
Start date:
15/12/2013
Due date:
15/12/2013
% Done:

100%

Estimated time:

Description

Check specifications with available sources.


Files

ACK7-2.jpg (37.4 KB) ACK7-2.jpg tin, 22/04/2015 23:17
ACK7-3.jpg (32.9 KB) ACK7-3.jpg tin, 22/04/2015 23:17
ACK7-5.jpg (34.5 KB) ACK7-5.jpg tin, 22/04/2015 23:17

History

#1

Updated by tin almost 6 years ago

  • Status changed from New to In Progress

AC input test, self-test result

AC source: Chroma 61504

110VAC, 60Hz - PASS
230VAC, 50Hz - FAIL2

90VAC 132VAC 180VAC 250VAC
50Hz FAIL2 FAIL2 FAIL2 FAIL2
60Hz PASS PASS PASS PASS
400Hz FAIL1 FAIL2 FAIL1 FAIL2

FAIL1 - Errors 200.1,200.2,200.6,200.7,201.2,401.1
FAIL2 - Errors 200.1,200.6,200.7,201.2

EDC MV106 1.00000V with REL set measurement overnight
Vmin = 1.0000001VDC, Vmax 1.0000042VDC

#2

Updated by tin almost 6 years ago

  • Status changed from In Progress to Closed
  • % Done changed from 0 to 100

Calibrated by Tek/Kei

Standards used:

F5700A + 5725A, Kei 2001-758 1Gohm standard, Agilent 33220A

#3

Updated by tin almost 6 years ago

AC input test, self-test result after cal

AC source: Chroma 61504

110VAC, 60Hz - PASS
230VAC, 50Hz - FAIL2

90VAC 132VAC 180VAC 250VAC
50Hz PASS PASS PASS PASS
60Hz PASS PASS PASS PASS
400Hz FAIL3 PASS FAIL3 PASS

FAIL3 - Errors 200.2,200.3,401.1

EDC MV106 1.00000V with REL set measurement overnight

#4

Updated by tin over 5 years ago

  • Target version changed from Keithley 2001 #5 to Keithley 2001 #7
#5

Updated by tin over 4 years ago