Project

General

Profile

Bug #855

Fix error 403.1 – Signal switching; ground switch

Added by tin almost 6 years ago. Updated over 4 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Target version:
Start date:
11/12/2013
Due date:
21/12/2013
% Done:

50%

Estimated time:

Description

Failure: Cannot measure 0.001V ±0.005V at A/D IN.

Common at pin 9 of U511 is multiplexed to pin 8 (OUT) and measured in the same
manner as the previous 400 series tests. Measure 0V at A/D IN.

History

#1

Updated by tin almost 6 years ago

  • Status changed from New to Resolved
  • % Done changed from 0 to 50

Fixed itself during probing.

#2

Updated by tin over 5 years ago

  • Target version changed from Keithley 2001 #7 to Keithley 2001 #6
#3

Updated by tin over 4 years ago

  • Status changed from Resolved to Closed

Also available in: Atom PDF