Support #1123
Fluke 52 II thermometer calibration - 89020093
Start date:
06/10/2014
Due date:
06/10/2014
% Done:
100%
Estimated time:
Description
5. Press ENTER for CAL A-d. Meter prompts 80. Set EDC MV106 to 80.0000mV.
EDC MV106, range 100mV | Kei2001 | DUT |
+80.0000mV | 80.00271mV | |
-8.00000mV range 10mV | -7.99550mV | -7997.0 |
6. CAL rEF
10. CAL donE - ENTER to store calibration.
Updated by tin over 8 years ago
- Status changed from In Progress to Closed
- % Done changed from 50 to 100